3. Apply Control Solution.
Shake the control solution vial thoroughly before
use. Squeeze out a drop and wipe it off, then
squeeze another drop and place it on the tip of the
vial cap.
Hold the meter to move the absorbent hole of the
test strip to touch the drop. Once the conrmation
window lls completely, the meter will begin
counting down.
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
2. Press the M button to mark this test as a control solution test.
If you press the M button again, the “QC” will disappear and this test
is no longer a control solution test.