Important Control Solution Information
Use only UltraTRAK PRO control solutions.
Check the expiration date on the control solution vial.
Do not use if expired.
Control solution, meter, and test strips should come to
room temperature 68°F to 77°F(20°C to 25°C) before
testing.
Shake the vial, discard the rst drop of control solution,
and wipe off the dispenser tip to ensure a good sample
and an accurate result.
Use only for 90 days after rst opening. Record the
discard date (date opened plus 90 days) on the control
solution vial. Discard after 90 days.
Store the control solution tightly closed at temperatures
36°F to 86°F (2°C to 30°C). Do not freeze.
The control solution range printed on the test strip vial is for UltraTRAK
PRO control solution only. It is used to test meter and test strip perfor-
mance. It is not recommended range for your blood glucose level.
PLEASE NOTE!